Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name Chiranjib
Log entry time 14:31:42 on January 24, 2009
Entry number 257596
This entry is a followup to: 257595
keyword=Diode laser 3 tripping
I checked the laser diode temperature settings . The min is 10 C and the
max is 50 C. So the limits are definitely not reached. I restarted the
temperature controller and the laser is ON now. But the gui showed that
the temperature controller was disabled while the other two lasers had
those ON. So we turned that back to ENABLE state. So far looks okay. It
seems that that was the problem and if it is , that is solved.
A copy of this log entry has been emailed to: yqiang