    User name A. Tobias

    Log entry time 08:17:11 on February 11, 2009

    Entry number 260463

    Followups:

    keyword=Target: NMR sweep for Trans. 3He not in correct frequency range?

    When taking one NMR sweep at the end of Owl Shift, we got only part of a signal? It looks like one of the coils current (DDM45_3disp1_read) slowly drifted during the shift and now the 3He resonance is higher in frequency. To avoid doing more NMRs, I decided not to do another sweep for now.

    A copy of this log entry has been emailed to: yqiang@jlab.org