keyword=Re :NMR screen (alarm is cleared)
It appears the main NMR code has frozen in certain subroutine at
finishing of last sweep. It either never reached the stand-by loop or
frozen in the loop (which is unlikely). Therefore the alarm sign has
not been cleared.
Abort and Restarting the NMR code solve all the problem.
It's still suggested to add a heart beat into the NMR code. So that, at
least when the code goes idle, people could notice that.
A copy of this log entry has been emailed to: yqiang