User name A. Camsonne
Log entry time 19:19:02 on February 27, 2009
Entry number 263735
keyword=Compton electron detector run list
It seems the situation is worse at 5 pass than at 4 passes, detector draw
about 1 uA in plan A at position 23000 compared to 26000 at 4 passes.
Might try a low current run later but the detector cannot be put in beam
for now. For reference lower limit switch is at ~34000 and 400 turns is
rougly 1.5 mm so detector reaches 1 uA at about 3 cm from the beam.
Though I don't have a good idea of how much the detectors are supposed to
draw in beam. Will have an idea before putting it in beam.
Position 22000 coincidence 2 planes, or 5 and width 11
Threshold
Run 16103 7
Run 16104 8
Run 16105 9
Run 16106 10 - 6 Hz no beam - 80 to 100 Hz with beam
Disable Plan A and strip 64 to 80 on plan C
Run 16107 10 - 1 Hz no beam - 50 Hz with beam
Position 22500
Run 16108 10 - 75 Hz with beam - 1 Hz no beam
Same condition single plane
Run 16109 10 - 1KHz with beam
Same condition - pos 21000
16110 10 - 1 KHZ
16111 30 - 50 Hz with beam
16112 25 - 134 Hz