User name R. Michaels
Log entry time 16:37:39 on April21,2009
Entry number 267245
Followups:
keyword=trig. mods, other problems
1. On R-HRS we set the MLU to "pulse" mode. This made the output
width normal. Looking through history this seems to be the
correct mode.
Bo recalls the L-HRS MLU does not have a strobe input. I didn't
climb over there to check, but it should be checked.
2. On R-HRS the inputs of trigger to the scaler are dangerous:
a) T1 is only 25 nsec wide. I think the min. is 10 nsec and
then the scaler doesn't count. Recommend >= 50 nsec.
b) Worse, the signals oscillate. Thus they may cause multiple
pulsing and artificial (wrong) deadtime. Suggest to clean up
these signals, like they are being fanned out wrong or a bad cable.
Bo is working on it.
Inputs to the normalization scaler should be clean boxes.