• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name V. Sulkosky

    Log entry time 22:22:17 on April 23, 2009

    Entry number 267667

    keyword=R-HRS Busy Checked and Contected to TS

    Bo and I checked the busy signals at the trigger supervisor by looking at
    them one by one. The signals from ROCs 1, 2 and 8 were about 11 usec. We
    then connected the BSY signal to the FB input on the TS.

    I attempted to look at the busy from the TS. Bob had suggested looking at
    this from the TIR module, but I couldn't see a signal. I noticed there
    was a 'BSY' from the TS2 module. Here the signal was about 90-100 usecs.

    I didnt' see any synch errors while running the DAQ in RightHRS
    configuration,
    which should be buffered and include the neutron detector.

    A copy of this log entry has been emailed to: rom.brads,moffit,zhaobo,yawei