• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name V. Sulkosky

    Log entry time 02:10:50 on April 24, 2009

    Entry number 267722

    Followups:

    keyword=R-HRS VDC Plots

    Attached are plots from the R-HRS VDC replay. For the working wires, the
    T0 looks good for all chambers. Also the number of wires hit per event
    is normal around five wires. The efficiency of the chambers looks
    reasonable for all the chambers but V1, which has the higher wire numbers
    drop to 90% efficiency.

    As far as the hit wires goes, the distribution is interesting. This
    distribution would lead me to believe that we have something in the middle
    of the acceptance for the R-HRS?

    A copy of this log entry has been emailed to: brads,jiang,jpchen,bogdanw,doug



    VDC TDC



    VDC Hits



    VDC Efficiency



    VDC Wire Hits