Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name V. Sulkosky
Log entry time 02:10:50 on April 24, 2009
Entry number 267722
Followups:
keyword=R-HRS VDC Plots
Attached are plots from the R-HRS VDC replay. For the working wires, the
T0 looks good for all chambers. Also the number of wires hit per event
is normal around five wires. The efficiency of the chambers looks
reasonable for all the chambers but V1, which has the higher wire numbers
drop to 90% efficiency.
As far as the hit wires goes, the distribution is interesting. This
distribution would lead me to believe that we have something in the middle
of the acceptance for the R-HRS?
A copy of this log entry has been emailed to: brads,jiang,jpchen,bogdanw,doug
VDC TDC

VDC Hits

VDC Efficiency

VDC Wire Hits