Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name Jin Huang
Log entry time 08:29:35 on April 24, 2009
Entry number 267800
keyword=Ref Cell Center Scan
A reference cell center search is performed with 2uA beam. Ref cell is
filled with 0 psig H2 to show a clearer rate change if beam hits on cell
wall, as well as carry away the beam heat from the wall.
center BPM on ref cell are
BPMA = (-1.0 2.1)
BPMA = (-0.5 2.1)
-----------------
Scan history
Raster BPMA BPMB T1 T3 Spot++ run
4x4 -1.3 1.5 -0.8 1.5 1.3K 1.8K 1313 No hit
6x6 -1.3 1.5 -0.8 1.5 1.3K 1.8K 1314 No hit
14x14 -1.3 1.5 -0.8 1.5 19K 12K ----
10x10 -1.3 1.5 -0.8 1.5 2.1K 2.7K 1316 Hit on one side
10x10 -0.8 1.5 -0.3 1.5 2.1K 3.0K 1317 Hit on one side
10x10 -0.8 1.0 -0.3 1.0 3.0K 4.3K 1318 Hit on one side, badly
10x10 -0.8 2.0 -0.3 2.0 1.8K 2.5K 1319 Hit on all four corners
10x10 -1.0 2.0 -0.5 2.0 1.9K 2.6K 1320 Hit on all four corners
10x10 -1.0 2.2 -0.5 2.2 1.3K 2.7K 1321 Hit on all four corners
A copy of this log entry has been emailed to: brads