    User name Yi/Chiranjib/Yawei

    Log entry time 18:34:00 on April 25, 2009

    Entry number 268416

    keyword=Laser interlock broken

    At around 4:30pm, all three COMET lasers were disabled due to "interlock disabled". Chiranjib and Yawei then checked the laser room and found out the interlock is actually enabled. Then by tracing the interlock voltage level, we found out that the first interlock chassis to which the three COMETs are connected is broken. So we re-route the interlock cable to bypass this chassis and connect all the COMET interlock cables to the second interlock chassis. Now the problem is solved and we enabled the laser to continue the pumping.

    A copy of this log entry has been emailed to: jpchen