    User name Yi Qiang

    Log entry time 22:26:46 on April 28, 2009

    Entry number 269551

    This entry is a followup to: 269546

    keyword=Third EPR with Anti-Parallel

    I flip the pumping direction to anti-parallel to the field.
    Unfortunately, I didn't see a clear Rb signal so I couldn't performe a
    Rb EPR as I did in parallel case. The result is consistent with the K
    EPR from parallel pumping.

    == Test Condition ==
    Cell: Dominic
    Oven Temperature: 230C
    Field Direction: Vertical -
    Pumping Direction: Parallel
    Laser Power: 3x25 COMET
    Cell Density: 7.830 amg

    == EPR of K ==
    Data directory: C:\AyeD\EPR\Data\EPR_AFP_20090428_2218
    Calibration Constant: 3.84 %/mV
    Polarization: 83.24%
    Field: 24.826 G
    EPR Frequency Shift: 147.84 kHz


    A copy of this log entry has been emailed to: jpchen, tdaver@wm.edu, yqiang, chiran, jinhuang



    Figure 1



    Figure 2



    Figure 3