    User name Yi Qiang

    Log entry time 23:14:53 on May 06, 2009

    Entry number 271337

    Followups:

    keyword=R-HRS is running with Unbuffered Mode and Neutron Detector ON

    Because that the busy signal from ROC8 still gives us problem, Vince and
    Bo unplugged it. So it leaves us two options, run R-HRS DAQ with
    unbuffered mode and Neutron Detector On or run R-HRS DAQ with buffered
    mode and ND OFF. We first tried to run unbuffered mode with ND ON. With
    10 uA, the online DATAMON shows a 16-17% dead time, and the offline
    NormAna analysis shows a 14 % dead time. Compared to the 5% dead time
    with buffered mode, Vince, Bo and I all think it's not bad so we will ask
    the shift crew to keep running production like this. We will try to find
    another opportunity to try to fix this problem.

    A copy of this log entry has been emailed to: tdaver@wm.edu, jpchen, jiang, doug, vasulk, zhaobo