Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name Ramesh/Xiaochao
Log entry time 17:19:40 on June15,2009
Entry number 281600
keyword=pvdis access activities (late entry)
Replaced the 100Hz tagger in the detector hut by a gate generator from the counting house middle room (also 100Hz), so we can control the tagger remotely.
Checked on the scopy if there is any loss in the tagger+signal coincidence w.r.t. the original tagger. The 100Hz tagger is too unstable for this scope check (we are looking for a 0.02% loss). When using a 10KHz pulser, did not see obvious loss (loss is <0.001kHz out of 10kHz).
Checked the 740 output (TS and PS) with and without taggers. It could drift up to 5mV.