keyword=PVDIS crate timing issue resolved
Bob, Kai, Ramesh ...
We do a few runs changing integration time in Green Monster
to check the PVDIS scalers. Recall, one bit is 2.5 usec.
Run ..... Integ. Time ..... Result
11451 .... 12000 ....... scalers good
11452 .... 13000 ....... scalers good
11453 .... 13300 ....... scalers fail intermittently (*)
11454 .... 13200 ....... scalers good
11455 .... 13300 ....... fail
11456 .... 13200 ....... good
(*) The scaler failure means that there is not exactly 1 event
in the FIFO. Could be zero or 2, etc. This is because of wrong
timing, see fig 1. This has been understood on a scope.
The bottom line is we need the integ time to be not more than
13200 in Green Monster !