Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name D. Parno
Log entry time 20:37:21 on August23,2009
Entry number 285555
This entry is a followup to: 285503
keyword=Right VDC Level Dropping
The right VDC discriminator level is continuing to drop. The set point is 4.00 but the values are now
Right Top: 2.12
Right Bottom: 3.84
Left Top: 3.89
Left Bottom: 3.96
Three and a half hours ago, the right top level was at 2.95V. The right top level now occasionally briefly drops below 2.0 V, which sets off alarms. The RC has paged the tech on call.