Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name paschke
Log entry time 01:27:39 on August24,2009
Entry number 285620
keyword=Right/Left HRS triggers at -2%
Comparing -2% kinematics, in focal place x vs. y and dp, between the right and left arms. Fig 1 is the left arm plot seen before, but with color contours. The left column in the T1 (happex detector) trigger and the right column is the T2 (S0) trigger.
Fig.2 is the corresponding right arm. Left column is the T2 (S0) trigger, and right column is the lead detector trigger scintillator (T3), which isn't very useful here.
But the comparison of T2 between the two arms is striking. Either the left arm trigger has a strong suppression of the high momentum side, or the right arm has a strong supression of the low momentum side.
FIGURE 1
FIGURE 2