• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name A. Camsonne

    Log entry time 19:56:51 on September 1,2009

    Entry number 287374

    keyword=Compton electron detector inhibit logic

    I added back 250 ns of dead time after an electron trigger to avoid additionnal triggers before the inhibit is generated. This seemed to crash the DAQ at high rate before. I managed to run at about 50 KHz rate since. Though still get the qid error when going to higher rates. I will double check what is exactly this error in adcHandler, this seems to happen less when the fifo eMaxDepth is made smaller, though dead time increases too when reducing this value so need to find a good balance I will try to optimize those values after the detector is checked out.