Log entry time 16:55:52 on September 2,2009
Entry number 287530
Followups:
keyword=Changed HP heater parameter to reduce false alarm
When running 100 uA, cryotarget alarms every time when beam comes on and off. I checked that it is due to the beam compenpesation slightly over-compensate the beam power, which causes the HP heater power goes to over 1000 when beam comes off and goes to zero when beam comes on. I changed one parameter dE/dx from 4.8 to 4.0, it reduces the HP heater to be less than 1000 when beam off. Seems working that it reduces the false alarm.