    Log entry time 16:18:54 on September 3,2009

    Entry number 287686

    keyword=raster current readout restored

    Raster current pickoff was restored on both HRS DAQs.

    To get the L-HRS picture, run "spot_L" or simply "spot"
    (spot is linked to L-HRS).

    To get R-RHS run "spot_R".
    (Yes, they both work now. I fixed that too.)

    The plots shown in fig 1 pop up automatically, in addition to
    the usual BPM info.

    This test was done with raster set at 11 x 8.5, our nominal max.
    The ADC should not saturate even if we go a little higher.
    However, this is the maximum attenuation I can achieve without
    reducing the ADC gate.

    What was wrong ?
    - Blown fuse on offset/atten board (board swapped out, meanwhile
    Jack replaced the fuse, so we have another spare).
    - One cable unplugged -- it interferes with Cavity electronics
    so I imagine someone unplugged it and forgot.