User name R. Michaels
Log entry time 21:22:54 on September 3,2009
Entry number 287765
Followups:
keyword=DAQ problem (fixed)
The HAPPEX DAQ stopped running because the R-HRS had no MPS.
Symptom is high deadtime, very low rate, and garbage data.
This turned out to be because of a bad chan 9 of the NIM/ECL
in CAMAC crate 1 slot 17. I moved to another NIM/ECL unit.
I also found the MPS cabling to be very messy and confusing.
A NIM/ECL was used as a fanout. Several copies went nowhere
(possible reflection). I've tried to clean this up but it may
need more cleanup. I think PVDIS is still getting it's copy.
Ran the 4 popular DAQ configs (1st 4 in list) and called it a night.
I didn't get to do the data distribution change because of this
problem. Maybe another day.