• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name paschke

    Log entry time 20:13:09 on September 6,2009

    Entry number 288323

    keyword=Dithing looks good, to be turned on

    Ran dithering again tonight, and the response looks significantly better.  This 
    is both because of some unexplained drift that may have improved the response 
    of the coils I tested yesterday, and because I ran additional coils that I 
    didn't test yesterday (I wasn't sure they were still enabled).  
    
    Results are below, from run #12055.  Everything was run with an amplitude of 
    100, except for coils 4 and 5 (amplitude = 250) and energy (amplitude = 50). 
    
    The absolute calibration of the energy change is in question.  According to 
    accelerators "FOPT" diagnostic, the dispersion at 1C12 is the same as it is in 
    the arc, but the arc should be ~2m and the 1C12 should be about 4 meters.  I 
    think we will need to get this settled eventually, since it does bear on our 
    ability to measure energy slopes. 
    
    I will implement this for regular running now.  Alexandre will be a little 
    cautious with the Compton electron detector to make sure there is nothing 
    dangerous there.
    
    Figure 1 is the dithering reaction in each bpm (there is a significant slow 
    drift over the measurement period).  Figure 2 is the same thing, but with both 
    dimensions read out for bpm12 and with 4a/4b decomposed into angle and position 
    on target.
    
    
    


    FIGURE 1

    FIGURE 2