    Log entry time 22:49:41 on September 24, 2009

    Entry number 291488

    Followups:

    keyword=Changed Right HRS/Compton BCM

    The FADC Compton BCM has not been working at 100uA. It turns out this is because it was using the U3 BCM which also goes to the right HRS (port #12 in the patch panel in the counting house; in the hall these signals are connected by a T). I changed this signal temporarily to the U1 BCM so that the Compton BCM will be okay until an access, where I will then be able to change the cabling in the hall. So the right HRS U3 BCM is U1 right now.