User name W. Deconinck
Log entry time 20:07:40 on October 23, 2009
Entry number 296359
Followups:
keyword=Overview of work performed in the hall earlier (with Ed, Alex)
While in the hall we replace the He level units on the left arm Q2 and
the right arm Q2 and Q3. When power cycling they gave repeatedly
different values (off by 15%) and different from values on a known good
unit (off by 60% in one case). The new units gave reasonable values of
80% for all levels. The suspicion is that radiation damage (units
located in the crawl space underneath the detector hut, open to lots of
particle spraying) caused the units to become unstable and the power
failure then gave them the last blow.
After replacing the units the spectrometer seemed to come up fine, but
unfortunately tripped soon afterward (see other entry, sorry that they
are out of order, I am trying to keep up).