Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name leacock
Log entry time 05:24:27 on March30,2010
Entry number 314699
Followups:
keyword=bpm saturation, bpm recovery
run 2402 shows the target bpms response to a current increase from 0 -> 50uA.
The target bpms adjust to the current increase in ~100 seconds. It takes bpm12
twice as long ~200 seconds to recover.
FIGURE 1
FIGURE 2