    keyword=heater and vacuum status

    Here's what Heidi told me. At about 9:15 she found the heater
    was unplugged, someone must have kicked the cord. See halog 325684.
    After plugging in at 09:30 in the graph the vacuum initially
    improved (the dip) then skyrocketed at 09:42. Then they applied
    vac-seal and the vacuum started coming down (after the big spike).

    They were limited in time (ALARA / RADCON) so they could not
    apply silicone, which would be the preferred fix but takes more time.

    I've confirmed that the heater outlet is drawing power now.