    User name R. Michaels

    Log entry time 11:25:23 on June 28, 2010

    Entry number 328901

    This entry is a followup to: 328895

    keyword=cleanup of 1024 Hz pulse

    The 1024 Hz pulse emerging from the RG58 cable from the R-HRS looked
    ratty, it could make a double-pulse.  I passed it through a discriminator
    and this cleaned it up.  I think nearly all these RG58 cables need to
    have their connectors replaced.  Meanwhile, the rates can be corrected
    easily offline; the scalers are inserted in the datastream every 4 sec.
    
    At the moment the 1024 Hz clock is normal again.
    


    A copy of this log entry has been emailed to: camsonne,vasulk