• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name A. Camsonne

    Log entry time 17:33:01 on September 04, 2010

    Entry number 332045

    keyword=DVCS DC current ADC setup

    The ADCs were set to differential ( Jumpers on R1 to R32 and no J1 and J2
    jumpers )
    
    The board to panel correspondance :
    First ADC2 VME adress 0x2
    Second ADC3   VME address 0x3
    Third  ADC 4  VME address 0x7
    Fourth  ADC1  VME address 0x9
    
    On the calorimeter panel, the display is in the same order as ADCs, 64
    channels of ADC1, then 64 channels of ADC2 and so on.
    
    Quickly tested with a 1.5 V battery it seems most channels are ok ( we
    did not test channel 32 since the last pin can have some different
    grounding) and calibration seems good too with 1.4 V read back.