User name R. Michaels
Log entry time 13:15:05 on September18,2010
Entry number 332469
keyword=DAQ fixed, was missing a gate
There was a report that "all histograms are empty" for L-HRS.
This was because a cable got knocked out upstairs, and so the
ADC in slot 24 of ROC4 was not getting a gate. When that happens
no data are read out (and I like it that way because then you
notice). We noted that for the ADCs in this crate, the ones
in slot 22 and 25 are getting back-panel gate and the ones in
23 and 24 are front-panel. Not sure if this is what DVCS wants.
I also was not 100% sure where the slot 24 gate came from; it was
grey twisted-pair cable #4 and I think I figure out where it had
been plugged in by Bogdan's crew (APEX remnant).