User name A. Camsonne
Log entry time 01:46:07 on October 10, 2010
Entry number 334165
keyword=DVCS DAQ status
-top crate mostly stable hardly an crash by itself
-firmware E06, timing is reasonably stable though channel 0 is missing on
all boards
-bottom crate crashes quickly - removed trigger interface to see if it
comes from there
-Magali looked at the logic module it seems to get S2m trigger but we
still have issues with the prescales, work around is to load a firmware
with hardcode prescales, we can do that will the FPGA programmers in place.
-no progress on the trigger, Magali will need a few hours to put a pulser
signal on each channel to check some timing, so this will most likely
happen on Monday during the injector work
Tomorrow we will test more the two crates setup.
We might be able to calibrate the calorimeter using ARS triggered by HRS
Electronics might be in reasonable shape for calibration but still needs
a lot of check before production : efficiency, synchronization and
consistency of our trigger with HRS T1 or T3 triggers