Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name R. Michaels / J. Roche
Log entry time 11:20:36 on November 4,2010
Entry number 338444
Followups:
keyword=helicity checks, part I
We checked the helicity in the Left HRS. Here's what we found and
what we did
1. Signals look good on a scope. Correct pattern (by eye) and
correct timing. No double-pulsing. Of course we cannot check
the pseudorandom pattern by eye, but the quad structure is there.
2. Swapped the scaler. This did not help.
At the moment we are stumped. There is a rare problem of helicity
in the scaler. I think the thing to try next is to run the DAQ at
the helicity rate (which will be 29 Hz this afternoon) and see if
the problem occurs. Then check the TIR data for helicity. If the
problem also exists in TIR, then there is a problem with the signals.
If the problem does not exit in the TIR then there's a problem with
the scaler or related software.