Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name R. Michaels
Log entry time 14:31:31 on November 8,2010
Entry number 339136
keyword=updated deadtime calculation
In the appendix below, I state what HAD been the deadtime
calculation prior to just now.
Here I state what is the new deadtime calculation. This affects
both datamon and the halog end-run (see fig 1).
DT = 1 - (T5 / (T10 + T7))
Appendix :
What was calculated before today:
Datamon: DT = 1 - (T5/T10)
Halog end-run DT = (T10-T5) - 1
They had the same sign but different denominator, and both
ignored T7. My understanding is that T7 is relatively new
and cannot be prescaled by the Trig. Super.
FIGURE 1