    keyword=Changed LH2 fan from 24Hz to 46 Hz and LD2 from 60 to 34 Hz.

    When we started the shift, we are running on LH2, but fan frequencies were still in the configuration for LD2 running: LH2 fan=24 Hz and LD2=60 Hz. Stopped run 9579.
    Changed loop 2 fan to 46 Hz (19%). started new run 9580.
    Changed Loop 1 fan to 34 Hz (14%).
    Adjusted J-T valves accordingly.