    keyword=Loop 1 temperature readout problem fixed

    Loop 1 temperature sensors take turn to read maximum every a few seconds. The ITC_502 readings are stable (ch 1=119.99, ch 2= 22,
    ch3= 22.27). Without a proper temperature reading, HP heater PID does not work. IOC reboot did not help.
    I first had it setup for feedback on the CERNOX_HX (24.5K).
    When given an access to the hall, I powercycled the ITC_502 and then did IOC reboot. It worked. I changed back to feedback on the 2nd CERNOX. All back to normal.