User name V. Sulkosky
Log entry time 20:57:44 on February 04, 2011
Entry number 346070
Followups:
keyword=Triggers Sent to Weldment
I connected T1, T2 and T8 from the right arm to the weldment via the
320ns RG-58 cables. Eventually T1 and T2 will be moved to RG-8 cables
once those are available. After arriving at the weldment patch panel, I
placed these triggers into a PS 707 discriminator to reshape the
signals. These signals were then connected to a level translater and
are then sent to the BigBite trigger supervisor. I will need to test to
make sure the TS sees these correctly.
David also sent a copy of T3 and T4 from the left HRS to the weldment via
the RG-8 cables to the weldment patch panel. I connected these the same
way as the right arm triggers. However the HV on the left arm was off,
so I was unable to verify the signals on a scope.
Eventually I will have to delay T3/T1 appropriately to form the
coincidence trigger T5.
A copy of this log entry has been emailed to: danez,yez,rom,lselvy