NEXT
Make New Entry,
Make Followup Entry
User name V. Sulkosky
Log entry time 15:51:34 on February 09, 2011
Entry number 346224
This entry is a followup to: 346098
keyword=Re: Right arm EDTM Setup
I rearranged the right arm EDTM pulser setup for simulated T1's to be
similar to the setup on the left arm. This entails using the back output
of the EDTM module for S1 on the right arm to feed the pulser into the
S2m electronics. Also the copy of the pulser from the weldment is now
coming up on one of the new RG-8 cables rather than the longer RG-58's.
Since the EDTM pulser is now coming earlier than before to the S2m
system, I was able to change the parameter "DRIGHT_S2M in the edtmsetup
file from 210 to 50. This should give us more headroom if we need it.
With the new setup, the "OR" of S1 EDTM now comes about 28 ns before the
"OR" of the S2m EDTM signals.
We also verified that the cosmic trigger has T1's timing defined off of
the S2m detector as it is on the left arm, i.e., S1 "OR" comes about
30 ns before the S2m "OR" for T1.
Now that both HRS EDTM systems have T1 and T3 setup correctly, we are
ready to simulate T5.
A copy of this log entry has been emailed to: danez,yez