S1 looks ok, the ribbon cable to the ADC had to be reseated, hence a few runs this afternoon without S1 amplitude. S1 TDC looks ok though double peak in S2m TDC are ok. Though looking closer one can see three distinct peaks. S2m ADC still look weird but I am wondering if it is not due to the trigger mostly being edtm. I cannot get the edtm TDC to remove the EDTM triggers. We will try to look at it with beam and maybe edtm off. S2m Right PMT ADC 7 looks different, either noisy pedestal or HV too low. VDC look ok Gas cerenkov looks ok A few missing channels on pion rejector even with pedestal subraction turned off. One can look at the replay of run 1178