Log entry time 18:37:29 on February22,2011
Entry number 348140
keyword=cryotargte back to operation mode with limited power for LD2
ESR has been stable for a while now.
Set cryotargte back to operational condition for LD2:
LD2 (loop1) at 22K, 21 psi, fan=48Hz, HP=~100watts,
LH2 (loop2) in rest mode: 19K, 31 psi, fan=24 Hz, HP=30 watts.
15K flow=15 g/s.
Without CHL supplement, 15 K supply temperatue increased from 13.5 to ~16K. With Hall C still take 35 g/s, we are at the limit.
This will limit us to have 100 watts in LD2, which only allows to run ~15 uA on 15 cm cell.
Need closely watching the cryo supply temrature, if it continously increases, we (A or C) will need to lower the heat load.