Entry number 349215
keyword=Update of HRS detectors and Create a new data base ~/DB/20110228/ due to move cables of S1-6R, S2-8R
It looks like all detectors in L-HRS are fixed. For R-HRS, only PS-18R is
still noisy. I traced the signal from upstairs patch panels and
downstairs pp, and they are clean. I changed its BNCtoECL black cable
from PP to ADC and also moved it to other ADC channel, but the broad
pedestal peak is still here. I really don't know which part has problem now.
I moved the cables of S1-6R,S2-8R at patch panel upstairs to the last two
inputs of S1-ADC group (after S0A,S0B), and I created a new data base
~/DB/20110228 to keep this change.
A copy of this log entry has been emailed to: danez@jlab.org