    keyword=RHRS: VDC Wires shoud multipeaks, as well as dp.

    I analyzed the last run#2060 and found that VDC wires has multi-peaks. We
    saw this problem when one HRS was detecting proton while S2m was making
    trigger, but currenlty RHRS is detecting electron. We look at dP, which
    also show multipeaks.
    
    Kalyan checked that there is not sieve slit in RHRS. I also posted a run
    with sieve in(#2022)
    


    A copy of this log entry has been emailed to: vasulk@jlab.org,doug@jlab.org



    Run 2060



    dP



    Run 2022 with Sieve Slot in