    I think BTA is fine.  From home ...
    
    What's actually not working is the script that writes to two EPICS
    variables which the BTA software then uses those variables to make
    the automated numbers.  It's not a show-stopper because the experimenter
    can (and should anyway) check & edit the BTA data.  As for what's wrong
    with the script: it seems that "caget hac_bcm_average" always results
    in zero, so it thinks the beam is off.  These data get logged to a log
    file and I can see that it was zero the whole time we've been running.
    So either this EPICS variable is wrong, or stuck at zero.  I suppose 
    we could substitute a scaler that's reading the BCM.
    
    The other job the script has to do is recognize if the DAQ is being used.
    This seems to be working.  It takes the "or" of the 2 HRS.  If either DAQ
    is being used, then we are using the beam.
    
    The script is a convenience feature.  Historically, it tends to be 
    reliable during steady production, but it's not very intelligent about
    various other uses (harp scan, Moller commissioning, etc) so one must
    always check.  And it seems to always break between experiments.
    
    
    