    User name jmaxwell

    Log entry time 09:15:16 on March 30, 2012

    Entry number 368326

    keyword=[Target] Main Flow Alarm Change

    The main flow has been a bit higher than usual as we're polarizing positive; the minor alarm at 55 has been tripping off pretty consistently. In addition to the high power we're seeing, this could also be an offset in the flow meter. I've set the minor alarm to 58, and left the major at 60, and that seems to have quieted the alarms.