The target nose fluid level is recovered now.
Future Target Operators: Be aware that the level probe appears to either skipping updates or capillary action in the tube is causing it to misread for indeterminant amounts of time. This could cause it to overfill or empty.
If the level reads -1 or 100.0 The nose has overfilled. -1 is the overlimit reading for the probe IT IS NOT EMPTY. You should put the run valve in manual mode, set it to 0.1 and wait for the level to come back down. Once the level is close to 50%, you can put the PID loop back on.
If you get a yellow LOW alarm due to a sudden drop in fluid, keep an eye on the level and be ready to use manual mode to recover the fluid.
If you get a red LOW alarm, immediately engage manual mode and set the run valve to 0.9 Watch for the nose to refill, and engage the PID loop once fluid level has recovered to around 50%.
Losing polarization while the fluid level is at 0% means the material is no longer under liquid, and you should immediately have MCC stop beam and you'll need to refill the nose and repolarize before resuming any data collection.
If at any point you are uncomfortable performing any of these procedures, call a target expert or the target on-call person.