    keyword=[Target] Polarization Versus Dose

    Here's my first look at the target polarization versus dose accumulated in 
    e/cm^2.  There may still be something little funny with the accumulation, 
    but it would probably be a linear scaling.
    
    The decay from last night is certainly slow, but not out of the realm of 
    possibilities.  Again, the raster is a bit bigger and the beam is shifted 
    to the right.  The first would slow the decay and the second would make us 
    think the decay is slowed.  
    
    We'll probably go ahead with an anneal tonight, even if the polarization 
    is still above 25%.  And it probably will be.
    