• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name R. Michaels

    Log entry time 07:46:19 on April15,2012

    Entry number 374802

    Followups:

    keyword=run number failed to increment

    This time on the R-HRS DAQ, the run number in the runcontrol GUI
    failed to increment. It shows 23386. But we are writing to run
    23387, as expected.

    Notice that the Output File in runcontrol says its run 23387.
    It looks like a casualty of the "too many databases" problem.
    One database for runcontrol and another for the rest of CODA.
    They can get out of synch.
    Doing a kcoda/startcoda should fix this.

    Fig 1 -- runcontrol GUI
    Fig 2 -- info about the files. they seem ok.

    After this run, we will "startcoda" for R-HRS.

    This is a different disease from earlier when we had duplicate
    files. But maybe it is related ?



    FIGURE 1

    FIGURE 2