• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT

    User name gilman

    Log entry time 18:00:57 on October 22,1998

    Entry number 3251

    keyword=helicity pulse counts

    PS on helicity setup:

    Previously I believe we sent the 100ns long logica output pulses to the scalers to be counted, but when going through the electronics I see we send the linear signals,
    that is the ~1 s long ones, to be counted. I do not know if any of the scalers
    have troubles counting ~1 Hz pulses, but it may be related to the extra counts we
    see on these. Will attempt to count the logical / short output pulses instead,
    when have the chance to set this up.