Main INDEX
Monthly INDEX
PREV
NEXT
User name Holmes
Log entry time 12:53:12 on January 22,1999
Entry number 9698
Followups:
keyword=Intensity asymmetry
HAPPEX DAQ gave a value last night and this morning of about 3000 ppm for the beam intensity asymmetry -- similar values were seen by Moller. In the past hour or so Matt Poelker and Jacque Ludwig have adjusted the Pockels cell HV offset to reduce this asymmetry by about a factor of 10.
This morning (HAPPEX run 8458):
Intensity asymmetry = -2624+-6 ppm width = 1059 ppm
x differences (BPM 3B) = 0.11 +- 0.05 microns width = 8.5 microns
y differences (BPM 3B) = 0.07 +- 0.02 microns width = 3.3 microns
Most recent (HAPPEX run 8466)
Intensity asymmetry = -296+-11 ppm width = 830 ppm
x differences (BPM 3B) = 0.11 +- 0.19 microns width = 15.0 microns
y differences (BPM 3B) = -0.14 +- 0.10 microns width = 3.3 microns