User name M. Liang
Log entry time 18:45:28 on January 25,1999
Entry number 10370
keyword=EPR/NMR calibration at 242.5 deg
Did two sets of EPR/NMR calibration for 242.5 deg target polarization angle. Did a FM scan first. EPR with 3 lasers on, NMR lasers off.
16:35 before moving to pick up coil position, EPR: ~35 kHz
16:42 moving to pick-up coil position, EPR: ~35 kHz
16:53 NMR: 9.77 mV
16:56 NMR: 9.74 mV
17:04 EPR: ~32 kHz
there seems polarization loss between the last two EPRs. Decided to do another set:
17:14 EPR: ~32 kHz
17:22 NMR: 9.21 mV
17:24 NMR: 9.18 mV
17:32 EPR: ~32 kHz
No polarization loss during this set observed. However, the polarization loss during the first set is confirmed by the NMR measurement in the second set. It's not clear where/how this happend. No correction of temperature change due to laser on/off is done yet, which may change the results. More detailed analysis will be done later. Plots in the plot-binder.