Main INDEX
Monthly INDEX
PREV
NEXT
User name gilman
Log entry time 00:37:46 on October 9,1999
Entry number 28008
This entry is a followup to: 28005
keyword=FPP front ch 1 U1,2,3 problem solved
We looked at the FPP gas and it was fine, all chambers bubbling normally.
We checked the LV anbd threshhold on the reeadout boards where there were problems, thinking some fuses mightbe bad, but all looked normal.
We looked on the scope at the output of the level shifter board and
saw signals from cosmics + noise on all the affected channels.
We pushed all the connectors back in nicely, and certainly we now see
in run 2173 oonline replay that all the low-efficiency channels look fine.
Suspect poor contact on some cable, and pushing on connectors solved it.