User name R. Michaels
Log entry time 15:52:18 on July14,2000
Entry number 45803
Followups:
keyword=trigger issues
Today we did the following for the trigger:
1. Checked that the pulser flag is registered with
good efficiency in the events. See end-of-run comment
for run 2896 where this was proven to be ok.
2. Added 16 nsec more delay to Aerogel so that it
nicely encompasses the S-Ray signal at MLU input.
These two inputs are patched through to the counting
room, to look on scope. The patches are in rack
CH01B04 (middle room ), near the bottom (a label
says "50 ohm to 1H75B03 in hall A"), and are
number 11 = S-ray, num 9 = Aerogel.
3. I also tried to look at TDC for Aerogel, but unfortunately
it reads 4095 both with and without beam; probably because
the signal is so wide and overlaps the end-of-range of TDC.