User name gilman
Log entry time 16:02:57 on November13,2000
Entry number 52017
keyword=day shift summary Nov 13, 2000
8:00 smooth shift changeover, but beam is off
run start time events comments
1270 8:29 - DAQ has communications problems, but stops
OK on second end-run attempt - there were
lots of communication failure to components msgs.
in CODA, but in components windows they all
responded ==> some intermittent problem??
1271 8:35 - coda responds ok as beam coming back
1272 8:47 1M beam erratic
1273 9:46 0.37 M end when MCC mentions problem
10:05 cavity problem, 10-15 minutes to fix... request access to check timing
~11:00 done with access
1274 10:12 43k "cosmic run"
1275 11:45 7k check raster with 5 uA
12:00 MCC calls to say there was trip and they are working on getting us beam back.
We has 5 uA from about 11:45 to 11:52; it tripped when we started going to 45 uA.
12:15 rad alarm problem, module needs to be replaced, 10 min - 1 hour estimate.
FPP HV alarmed, channel 8.1.9 was showing 0 V, reset arcnet and turned the
FPP check script back on.
1276 12:40 - raster check
1277 12:43 0.27M data! end for timing change - 32 ns taken out in ch 11
1278 13:02 1 M - good -
1279 14:03 1 M - good -
1280 15:03 1 M - good -
1281 16:00 started...