    User name adev

    Log entry time 23:02:39 on December 17, 2000

    Entry number 53762

    keyword=End_of_Run_1975
    ****************************
     Run Number: 1975
    End of Run Comment:
    Tar=LH2-15cm,Type=Elastic,Lmom=3.3593,Rmom=1.0,Ene=4.58,Angle_R=12.509,Angle_L=23.768
    
    
    Time of run = 55.510287 min
    Av. current (on 3x) = 45.660966 uA
    Av. current (on 1x) = 45.559304 uA
                   |   on 1x    |   on 3x    |   on 10x
    Av. charge (mC)|  151.7406  |  152.0792  |  150.8015
    up-down    (mC)|    2.1283  |    1.7760  |    2.8831
     Trigger   | Scaler/charge(3x)|Scaler/time/ps|  Scaler    | Prescale| Scaler/charge(1x)
        1      |   6410719.9 /mC  |    1.626 Hz  | 974937150  | 180000   |   6425025.0 /mC
        2      |    104501.7 /mC  |    0.954 Hz  |  15892532  |  5000   |    104734.9 /mC
        3      |      6395.3 /mC  |  292.015 Hz  |    972591  |     1   |      6409.6 /mC
        4      |    746487.9 /mC  |   17.043 Hz  | 113525280  |  2000   |    748153.6 /mC
        5      |       594.4 /mC  |    0.000 Hz  |     90399  | 65535   |       595.7 /mC
        output                       300.214 Hz
    Av. deadtime from accepted triggers = 0.0367
    Av. deadtime from gated scalers = 1.0000
     Trigger   |  Plus Helicity   |  Minus Helicity |  Sum 
        1      |  292676.19  Hz     |    292914.92 Hz   |  292719.67 Hz
        2      |  4771.2386  Hz     |    4784.8279 Hz   |  4771.6477 Hz
        3      |  292.57646  Hz     |    291.57495 Hz   |  292.01525 Hz
        4      |  34083.323  Hz     |    34108.377 Hz   |  34085.358 Hz
        5      |  27.204169  Hz     |    27.179424 Hz   |  27.141816 Hz
        output |  0.0  Hz   |    0.0 Hz |   300.214 Hz
    Accepted triggers  = 999899.00000
     Time of run = 55.510287 min
     Time of run h- = 27.441715 min
     Time of run h+ = 27.44126 min
     1-(sum of time hel + and -)/( run time) = 1.1301e-02
    Rate of Left Arm Pulser for electronic DT  = 10.168386
    Rate of Right Arm Pulser for electronic DT  = 5.0423087
    Rate of Left Arm Pulser&trig for electronic DT  = 10.117644
    Rate of Right Arm Pulser&trig for electronic DT  = 5.0423087
    Left Arm Electronic DT  = 0.0049901084
    Right Arm Electronic DT  = 0.0
    Rate of Left Strobe  = 57056.3
    Rate of Right Arm Strobe  = 0.0
       helicity    |  Gate on + and run  |  Gate on - and run  |  Gate on run 
          +        |  11.263678  Hz         |    0.0 Hz        |  11.263678 Hz
          -        |  0.0  Hz        |    11.263378 Hz       |  11.263678 Hz
     
        Signal     |  Gate on + and run  |  Gate on - and run  |  Gate on run  |
                   |     Charge   (C)    |     Charge   (C)    | Charge   (C)  
       upstream    |   0.0755157         |   0.0755754         | 0.1528047     
       upstream 3x |   0.0755925         |   0.0756510         | 0.1529672     
       upstream 10x|   0.0752402         |   0.0752937         | 0.1522431     
      downstream   |   0.0744640         |   0.0745228         | 0.1506764    
     downstream 3x |   0.0747136         |   0.0747702         | 0.1511912    
     downstream 10x|   0.0738142         |   0.0738656         | 0.1493600    
     
     
        Signal     |  Asymmetry
                   |           
       upstream    |   3.95125850e-04 
       upstream 3x |   3.86793480e-04 
       upstream 10x|   3.55401670e-04 
      downstream   |   3.94665840e-04 
     downstream 3x |   3.78636350e-04 
     downstream 10x|   3.48050310e-04 
     
    
    ****************************
    End-of-run EPICS data for Run Number 1975 at Sun Dec 17 23:02:26 EST 2000
    Right Arm D1 momentum                    : 1
    Right Arm arm angle                      : 12.5091
    Left Arm D1 momentum                      : 3.35931
    Left Arm arm angle                        : 23.7678
    Tiefenbach Hall A energy                : 4582.75
    Right Arm arm mispointing (HPT)          : -1.1932
    Left Arm arm mispointing (HPT)            : 5.3042
    Right Arm Q1 power supply current        : 752.01
    Right Arm Q1 momentum                    : 1.0001
    Right Arm Q1 gauss meter                 : -0.018768
    Right Arm Q1 power supply polarity       : 0
    Right Arm Q2 power supply current        : 430.497
    Right Arm Q2 momentum                    : 1.0003
    Right Arm Q2 gauss meter                 : 0.22556
    Right Arm Q2 power supply polarity       : 0
    Right Arm D1 power supply current        : 342.987
    Right Arm D1 momentum                    : 0.999998
    Right Arm D1 NMR                         : 0.371087
    Right Arm Q3 power supply current        : 397.833
    Right Arm Q3 momentum                    : 0.99981
    Right Arm Q3 gauss meter                 : 0.17781
    Right Arm Q3 power supply polarity       : 1
    Left Arm Q1 power supply current          : 2527.49
    Left Arm Q1 momentum                      : 3.35954
    Left Arm Q1 gauss meter                   : 0.21419
    Left Arm Q1 power supply polarity         : 1
    Left Arm Q2 power supply current          : 1446.51
    Left Arm Q2 momentum                      : 3.35929
    Left Arm Q2 gauss meter                   : -0.7614
    Left Arm Q2 power supply polarity         : 0
    Left Arm D1 power supply current          : 1192.67
    Left Arm D1 momentum                      : 3.3593
    Left Arm D1 NMR                           : 1.24417
    Left Arm Q3 power supply current          : 1337.58
    Left Arm Q3 momentum                      : 3.35975
    Left Arm Q3 gauss meter                   : 0.6671
    Left Arm Q3 power supply polarity         : 0
    Average current of the two cavities     : 58.0551
    Current on Unser monitor                : 51.6636
    Current on Upstream bcm                 : 57.2833
    Current on Downstream bcm               : 58.8269
    Upstream Cavity Coefficient             : 77
    Downstream Cavity Coefficient           : 78.9
     Current on IBC0L02 (source)            : 58.588
    MBSY1c Hall A Beam energy               : 4592.25
    MBSY1c magnet current (amp)             : 191.553
    MBSY1c Bdl                              : 1.146e+06
    MBSY3c Hall C Beam energy               : 2329.44
    Tiefenbach Hall A energy                : 4582.74
    Tiefenbach Hall C energy                : 2323.98
    Delta(E)/E width (MV_AI_DXWID)          : 2.12419
    IPM1C12.XPOS                            : -0.00600353
    IPM1H04A.XPOS                           : 0.399014
    IPM1H04A.YPOS                           : 0.929565
    IPM1H04B.XPOS                           : -0.0255971
    IPM1H04B.YPOS                           : 0.109901
    Right Arm Arm Collimator                 : 0.364014
    Left Arm Arm Collimator                   : 0.868506
     HacR_HPM_LVDT:VALr                     : -1369
     HacR_VPM_LVDT:VALr                     : -1669
     HacR_HPE1313A:0_Ch58                   : -0.569824
     HacR_HPE1313A:0_Ch55                   : 22.7539
     HacR_HPE1313A:0_Ch54                   : -0.842773
     HacR_alignROLL                         : -0.362525
     HacR_alignPITCH                        : 0.0123861
     HacR_alignHPT                          : -1.19152
     HacL_HPM_LVDT:VALr                     : -8123
     HacL_VPM_LVDT:VALr                     : 1359
     HacL_HPE1313A:0_Ch58                   : -0.129395
     HacL_HPE1313A:0_Ch55                   : 23.7793
     HacL_HPE1313A:0_Ch54                   : -0.283203
     HacL_alignROLL                         : 0.0817571
     HacL_alignPITCH                        : -0.129473
     HacL_alignHPT                          : 5.30514
    haBDSPOS                                 : 5.354e+06
    haL1_Fan_frequency_R                     : 0
    haL2_Fan_frequency_R                     : 63.5
    haL3_Fan_frequency_R                     : 0
    haH2_P_Fill_Target_R                     : 21.78
    haH2_P_Exhaust_R                         : 21.86
    haH2_P_Fill_Bottle_R                     : 22.21
    haH2_DP_Fill_Exhaust_R                   : 0
    haD2_P_Fill_Target_R                     : 0
    haD2_P_Exhaust_R                         : 0
    haD2_P_Fill_Bottle_R                     : 0
    haD2_DP_Fill_Exhaust_R                   : 27.665
    haSC200_6_Sensor_1_R                     : 242.45
    haSC200_6_Sensor_2_R                     : 17.07
    haSC200_6_Sensor_3_R                     : 3.04
    haSC200_6_Sensor_4_R                     : 40.13
    haHe_Omega_DPG_Sensor_R                  : -0.032
    haITC502_1_Sensor_1_R                    : 94.98
    haITC502_1_Sensor_2_R                    : 89.87
    haITC502_1_Sensor_3_R                    : 94.98
    haITC502_2_Sensor_1_R                    : 19
    haITC502_2_Sensor_2_R                    : 19
    haITC502_2_Sensor_3_R                    : 19.19
    haITC502_4_Sensor_1_R                    : 76.5
    haITC502_4_Sensor_2_R                    : 19.2
    haITC502_4_Sensor_3_R                    : 66.9
    haITC502_5_Sensor_1_R                    : 62.78
    haITC502_5_Sensor_2_R                    : 60.97
    haITC502_5_Sensor_3_R                    : 159.7
    haITC502_6_Sensor_1_R                    : 96.97
    haITC502_6_Sensor_2_R                    : 99.97
    haITC502_6_Sensor_3_R                    : 26.27
    Right Arm Arm Collimator                 : 0.364014
    Left Arm Arm Collimator                   : 0.868506
    Wien angle                               : -6.99926
    Helicity Mode ON/OFF Random/Toggle       : OFF
    Wien BDL                                 : -203.2
    Wien field                               : 5.6245
    Laser attenuation Hall A                 : 461
    Laser power  Hall A                      : 73
    Half-wave plate (off=out)                : ON